Development of the Ion Source of Glow Discharge/Mass Spectrometry for the determination of trace elements
미량원소 분석을 위한 GD/MS 이온원의 개발에 관한 연구
- Woo, Jin Chun (Inorg. Anal. Chem. Lab., Korea Standards Research Institute and Science) ;
- Lim, Heoung Bin (Inorg. Anal. Chem. Lab., Korea Standards Research Institute and Science) ;
- Moon, Dae Won (Inorg. Anal. Chem. Lab., Korea Standards Research Institute and Science) ;
- Lee, Kwang Woo (Inorg. Anal. Chem. Lab., Korea Standards Research Institute and Science) ;
- Kim, Hyo Jin (Department of Pharmacy, Dong Duck Women's University)
- 우진춘 (한국표준과학연구원 무기분석연구실) ;
- 임흥빈 (한국표준과학연구원 무기분석연구실) ;
- 문대원 (한국표준과학연구원 무기분석연구실) ;
- 이광우 (한국표준과학연구원 무기분석연구실) ;
- 김효진 (동덕여자대학교 약학대학)
- Received : 1992.03.16
- Published : 1992.06.25
Abstract
Analytical detection limits and Relative Ion Yield (RIY) by a jet type ion source glow discharge mass spectrometer(GD/MS) have been measured. With a jet type ion source, the sample loss rate for a Cu sample is 0.23 mg/min with 0.1 L/min gas flow rate and 0.11 mg/min with no gas flow rate. However, the ion intensity of Cu does not change significantly with thee variation of the gas flow rate. The RIY values obtained from the calibration curves of the six copper based standards were between 0.57 of Fe and 3.5 of Cr. The detection limits of most elements were in the range of 0.9 and 2.0 ppm when the glow discharge was operated at 4 mA, 1000V.
금속시료의 미량원소를 분석하기 위하여 jet형과 글로우방전 이온원을 제작하여 질량 분석기와 연결시킨 후 상대이온세기와 검출한계를 측정하였다. Jet형 이온원으로 구리시료의 시료 손실속도를 측정하였을 때 가스흐름속도가 0.01L/min이었을 때 0.23 mg/min이고 가스흐름이 없을 때는 0.11 mg/min이었다. 그러나 구리의 이온세기를 측정하였을 때는 두 이온원에서 큰 차이가 없었다. 여섯 개의 동합금시료의 검정곡선으로부터 구한 각 원소의 상대이온세기는 철의 0.57부터 크롬의 3.5 범위였다. Jet형 이온원으로 4mA, 1000V 방전시 순수 구리시료의 각 원소에 대한 검출한계는 0.9 ppm에서 2 ppm 수준으로 나타났다.