Determination of Trace Impurities in High Purity Aluminum by Instrumental Neutron Activation Analysis

고순도알루미늄의 비파괴 중성자방사화분석

  • Published : 1992.06.01

Abstract

Nondestructive neutron activation analysis of copper in high purity Aluminum samples which can be used as a parameter of impurity is investigated and determined. Other 23 trace impurity elements in the samples are also determined. In the analysis of copper, the new irradiation method using thermal column was applied to reduce the interfering activity of 24Na produced by 27Al(n,$\alpha$)24Na reaction induced by fast neutron. As a result, the interference can be reduced to 100 times more than other activation methods. Also the influence by activity of 24Na is found in the range of 2~3 %. It has been observed that the copper contents in so-ratted“six nine”class standard aluminum samples are about 0.54 $\pm$ 0.08 ppm. By the comparison with other values re-ported, our results are reasonable and can be available as a improved routine analysis.

고순도알루미늄중 불순물의 Parameter로 이용될수 있는 구리의 비파괴 방사화분석법의 고찰 및 23종의 극미량불순성분원소의 함량을 분석하였다. 즉 구리의 분석은 원자로의 속중성자에 의한 27Al(n,$\alpha$)24Na반응으로 생성되는 24Na의 방사능을 감소시키기 위하여 Thermal Column을 이용하였고 다른 조사공을 이용한 경우보다 약 100 배 정 도 방해 요인을 감소시킬 수 있었다. 24Na 에 의한 영향은 2-3 %범위 이하이었다. 이 방법에 의해 표준알루미늄(6 nine class)시료로부터 구리를 정량하였고 아울러 기타 불순원소들을 일상 방사화분석법에 의해 정량하였다. 구리의 함량은 0.54$\pm$0.08 ppm이었다. 이러한 결과는 문헌값과 비교할때 타당성이 있었고 일상분석에 이용할 수 있는 좋은 방법으로 여겨진다.