초록

We studied the characteristics of the compound electrode and found that the compound electrode provides the four-electrode method in a compact form. We developed a new method of measuring the skin impedance using simple electrodes at low frequencies. At high frequencies where the effect of internal tissue impedance is not negligible, we used the compensation method using compound electrodes since they measure the voltage right under the skin. At 50 kHz, we measured the real part of the skin impedance of less than $80{\Omega}$ on the thorax. We propose a simple instrument which can measure accurate skin impedance at various frequencies.

키워드

참고문헌

  1. IEEE Trans. Med. Imaging v.10 Iterative reconstruction methods using regularization and optimal current patterns in electrical impedance tomography R.Hua;E.J.Woo;J.G.Webster;W.J.Tompkins
  2. IEEE Trans. Biomed. Eng. v.BME-35 Skin impeldance from 1 Hz to 1MHz J.Rosell;J.Colominals;P.Riu;R.Pallas;J.G.Webster
  3. Medical instrumentation : application and design(2nd ed.) Measurement of flow and volume of blood J.G.Webster(ed.)
  4. Electrical Impedance Tomogrphy J.G.Webster(ed.)
  5. Med. & Biol. Eng. & Comp. v.14 Characteristics of skin admittance for dry electrodes and the measurement of skin moisturisation Y.Yamamoto;T.Yamamoto;T.Ozawa
  6. Med. Biol. Eng. v.14 Electrical properties of the epidermal stratum corneum T.Yamamoto;Y.Yamamoto
  7. Med. Biol. Eng. v.14 Dielectic constant and resistivity of epidermal stratum corneum T.Yamamoto;Y.Yamamoto