ML-6M선형 가속기에서의 BEAM 특성에 대한 고찰

  • 문언철 (원광대학교 의과대학 부속병원 치료방사선과) ;
  • 윤병운 (원광대학교 의과대학 부속병원 치료방사선과) ;
  • 오양환 (원광대학교 의과대학 부속병원 치료방사선과)
  • Published : 1992.11.01

Abstract

The beam characteristics and dosimetric measurements of the 6MV X-ray and 6MeV electron beam from a ML-6M linear accelerator are examined. The Percent Depth Dose(PDD) table and the tissue Maximum Ratio(TMR) table are taken from measurement as a function of the field size and the depth. The calculated TMR table from PDD table is compared with those from measurement. Other beam characteristics such as output factor, beam profile(including flatness, symmetry and penumbra), wedge, and the variation of Dmax are presented. All of these dosimetric measurements sufficiently characterized the beam to permit safe clinical use.