Development of a New Radiotherapy Technique using the Quasi-Conformation Method

Quasi-Conformation 치료를 위한 새로운 방사선치료기술의 개발

  • Choi, Tae-Jin (Department of Therapeutic Radiology, School of Medicine, Keimyung University) ;
  • Kim, Jin-Hee (Department of Therapeutic Radiology, School of Medicine, Keimyung University) ;
  • Kim, Ok-Bae (Department of Therapeutic Radiology, School of Medicine, Keimyung University)
  • 최태진 (계명대학교 의과대학 치료방사선과착교실) ;
  • 김진희 (계명대학교 의과대학 치료방사선과착교실) ;
  • 김옥배 (계명대학교 의과대학 치료방사선과착교실)
  • Published : 1991.12.01

Abstract

The quasi-conformation therapy was performed to get a homogeneous dose distributions for irregeular shaped tumor lesion by using the arc moving beam and beam modifying filter which was made by cerrobend alloy($\rho$=9.4 g/cc) metal. In our dose calcuation programme, it was fundmentally based on Clarkson's method to calcuate the irregular multi-step block field in rotation therapy. In this study, the expected relative depth doses under multipartial attenuator agree well with measured data at same plane. The results of comparison the dose computation with that of TLD measurement are very closed within ${\pm}5\%$ uncertainties in the irradiation to phantom with quasi-comformation method. And it has shown that irregular typed multi-step filter can be applied to quasi-conformation therapy in high energy radiation plannings.

종양모양에 거의 일치하는 선량분포를 얻기위해 다층부정형필터를 이용하는 방법을 보였으며, 필터의 재질은 Lipowitz (일명 cerrobend alloy) 금속체를 이용하였다. 선속내 필터의 놓임은 일차선의 감쇄와 함께 조직내 산란선 기여율의 변화가 예상되므로 필터의 두께에 따른 SMR의 변화가 계산에 이용되어야 된다. 이에 본 연구의 계산선량은 실측치의 $3\%$이내에서 잘 일치됨을 알 수 있었다. 다층부정형필터를 이용한 펜톰의 원호조사의 선량평 가는 동일 필터에 의한 고정조사의 선량분포가 적용되었으며, 가상종양에 거의 일치하는 선량분포를 얻을수 있었다. 한편, 컴퓨터모의계산된 선량과 열발광산량계(teflon-embedded TLD)의 실측선량은 비교적${\pm}5\%$ 범위내에서 잘 일치됨을 알 수 있었으며, 이 치료방법은 현재 차폐체로 많이 이용되고 있는 Lipowitz 금속체를 사용함으로써 보다 쉽게 응용되어질 것으로 생각된다.

Keywords