Journal of Yeungnam Medical Science
- Volume 7 Issue 2
- /
- Pages.115-120
- /
- 1990
- /
- 2799-8010(eISSN)
A Study on the dosimetry in boundary of shielding block in high energy irradiation
고에너지 방사선치료에서 차폐물 경계부위의 선량분포에 관한 고찰
- Kim, Myung-Se (Department of Therapeutic Radiology, College of Medicine, Yeungnam University) ;
- Kim, Sung-Kyu (Department of Therapeutic Radiology, College of Medicine, Yeungnam University) ;
- Shin, Sei-One (Department of Therapeutic Radiology, College of Medicine, Yeungnam University)
- Published : 1990.12.30
Abstract
Scatter-air ratios are used for the purpose of calculating scattered dose in the medium. The computation of the primary and the scattered dose separately is particularly useful in the dosimetry of irregular fields with shielding block in radiation field, dose distribution of scattered radiation using 18MeV Linear accelerator and Co-50 teletherapy measured. The effect of scattered radiation dose by protecting block was been ignored in radiation therapy, 2-3% of scattered radiation may be 90-200 cGy which could be influence vitial complications such as cataract, oligospermia or sterility. So that exect calculation of such scattered radiation especially for large field
18MeV 선형가속기와 코발트 원격치료기를 사용하여 방사선 조사면적 내에 차폐물이 있는 경우, 차폐물에 의한 산란선이 선량분포에 미치는 영향에 관하여 고찰하여 다음과 같은 결과를 얻었다. 1. 차폐물이 이루는 각이 예각일수록 산란선의 효과는 더 크게 나타났다. 2. 차폐물의 넓이가 좁을수록 산란선의 효과는 더 크게 나타났다. 3. 조사면적에 따른 출력특성은 선형적이지만, 차폐물에 의한 출력특성은 조사면적에 대하여 선형가속기는 거의 무관하게 나타났으며, 코발트 원격치료기는 기울기가 적은 선형성을 나타내었다.