EXPERIMENTAL STUDY OF INFLUENCE ON RADIOGRAPHIC DENSITY AND CONTRAST BY THE CHANGE OF KILOVOLTAGE AND EXPOSURE TIME

관전압과 X선노출시간의 변화가 X선사진의 흑화도와 대조도에 미치는 영향에 관한 실험적 연구

  • Lee Byeong Do (Department of Oral Radiology, College of Dentistry, Kyung Hee University) ;
  • Lee Sang Rae (Department of Oral Radiology, College of Dentistry, Kyung Hee University)
  • 이병도 (경희대학교 치과대학 치과방사선과학교실) ;
  • 이상래 (경희대학교 치과대학 치과방사선과학교실)
  • Published : 1990.06.01

Abstract

For the study of the influence of kilovoltage and exposure time on radiographic density and contrast, we measured radiographic density of aluminum step wedge which composed of contiguous 8 steps wedges of 2-16㎜ thickness with densitometer. Aluminum step wedge was radiographed on Kodak ultraspeed DF-58 and Ektaspeed EP-21 film with range of 60-90 kVp and 5-60 impulse and subject contrast of aluminum step wedge with constant radiographic density and image contrast percentage without radiographic density was evaluated. Then we evaluated the film quality of teeth and their surrounding structure according to the change of kVp and exposure time by score rating method. The obtained results were as follows: 1. Radiographic density was related to the change of kilovoltage, especially in increased exposure time. 2. With constant radiographic density, subject contrast of thin aluminum step wedges was greater in low kilovolt age than high kilovoltage, but kilovolt age had not great influence on subject contrast of thick aluminum step wedge. On the other hand, radiographic density difference between 2mm and 16mm aluminum step wedge was decreased according to in- creasing kilovoltage. 3. Without constant radiographic density, image contrast percentage was decreased with increasing kilovoltage, but was not related with the change of exposure time. 4. Radiographic contrast of teeth and their surrounding structure which was taken with the range of 60-90 kVp and 6-30 impulse had not great influence on film quality.

Keywords