Electron Dosimetry of Shaped Fields on Mevatron KD 67-7467

Mevatron KD 67-7467의 변형조사면에 대한 전자선 선량측정

  • U Hong (Department of Radiation Oncology Kyungpook National University College of Medicine) ;
  • Samuel Ryu (Department of Radiation Oncology Kyungpook National University College of Medicine) ;
  • H. D. Kang (Department of Physics Kyungpook National University College of Natural Sciences)
  • Published : 1990.10.01

Abstract

A method of making inserts for shaped fields in electron beam therapy on the Mevatron KD 67-7467 Linear Acclerator is introduced. The inserts are made from an alloy called Lipowitz metal. These are designed to fit the inside of the standard Siemens cones. Studies have shown that this method does not adversely affect field flatness. However, if the ratio of shaped field to open field is greater than about 70%, the output dose is significantly changed by the inserts. Because the cone ratios for the fields do not follow the open cone ratio curves on the Mevatron KD 67-7467, we separated the cone ratio suggested by Biggs into two parts, the insert ratio and the cone factor. The dosimetry for these shaped beams has been investigated extensively.

전자선 치료에 있어서 조사면을 부분적으로 차폐할 수 있는 차폐체 제작 방법을 제시하였다. Lipowitz 합금으로 만든 차폐체는 표준 Siemens cone에 적합하며, 조사면의 평탄도는 차폐체에 의한 영향을 거의 받지 않았다. 그러나 개방 조사면에 대한 차폐 조사면의 비가 0.7 이상이면 차폐체에 의한 선량의 차이가 있음을 알았다. Biggs 등이 제안한 Cone ratio는 차폐 조사면의 Cone ratio curve가 개방조사면의 Cone ratio를 따르지 않기 때문에 본 연구에서는 Cone ratio를 Cone factor와 Insert ratio로 나누어 차폐 조사면에 대한 선량 변화를 넓게 측정하였다.