A Glove Box for the Subsampling of Suboxic and Cold Core Sediment

아산화 및 저온의 코아 퇴적물 중 부시료 채취를 위한 글러브박스

  • Published : 1990.12.01

Abstract

To solve two problems, temperature shock to the man and difficulty in exchanging the air in the glove box into N2 gas, during the subsampling of Suboxic and cold core sediments, a portable glove box with a refrigerating machine and a gas exchanging bag was constructed for subsampling of sediment cores on board. The refrigerator of the glove box can cool down 200 litter air at 30$^{\circ}C$ to 2$^{\circ}C$${\pm}$2$^{\circ}C$ within 5 minutes. The box was successfully operated during the second KORDI's deep sea research cruise of 1989 in the North equatorial Pacific. Pore water data obtained from the cruise show no evidence for artifacts caused by warming up or oxidation of sediments during subsampling.

아산화 및 저온의 코아 퇴적물 중 부시료 채취시 실험자에게 미치는 온도 충격과 글러브박스내의 공기를 질소가스로 교체할 때의 난점을 개선하기 위하여, 이동이 가능 하고 냉장고와 기체교환풍성을 가진 글러브박스를 제작.실험했다. 본 글러브박스는 3$0^{\circ}C$의 실온에서 200미터의 기체를 5분 이내에 2$^{\circ}C$$\pm$2$^{\circ}C$로 냉각시킬 수 있다. 또한 기 체교환풍선을 이용해 글러브박스내의 기체를 질소가스로 쉽게 교체할 수 있다. 이 글 러브박스는 태평양의 클라리온, 클리퍼톤 균열대 지역에서 수행된 KODOS-89 탐사에서 성공적으로 작동되었으며, 얻어진 공극수 자료로부터 퇴적물 시료의 가열 또는 산화의 흔적을 찾을 수 없었다.

Keywords