Abstract
A carbon paste coated-wire ion-selective electrode for chloride (carbon chloride-CWE) was constructed using epoxy resin, ion-exchanger and carbon powder as a polymer membrane. Its utility, the composition of a polymer membrane, the response characteristics, and the selectivity were examined and applied to the environmental water analysis. The carbon chloride-CWE was prepared using a silver wire, which was covered with silver chloride and then coated with epoxy resin into which chloride ion-exchanger and carbon powder were previously incorporated in advance. The response of the carbon chloride-CWE was Nernstian for $1.0{\times}10^{-2}-2{\times}10^{-5}$ M chloride and the useful pH range from $10^{-2} M Cl- to 10^{-4} M Cl^-$ was 3.0-9.0. Furthermore, the selectivity of chloride over iodide, bromide, and cyanide was much improved compared with those for a solid state epoxy body chloride electrode and a liquid membrane chloride electrode. The carbon chloride-CWE was applied to determine Cl^-$ in tap and ground water. The obtained results were in good agreement with those by the established methods such as spectrophotometric or other chloride-selective electrode methods.