6MV X-선에 대한 삼차원적 조직보상체의 연구

A Study of Three-dimension Tissue Equivalent Compensator for 6MV X-Rays

  • 김옥배 (계명대학교 의과대학 치료방사선학과학교실) ;
  • 최태진 (계명대학교 의과대학 치료방사선학과학교실) ;
  • 서수지 (계명대학교 의과대학 치료방사선학과학교실)
  • Kim, Ok-Bae (Department of Radiology, Keimyung University School of Medicine) ;
  • Choi, Tae-Jin (Department of Radiology, Keimyung University School of Medicine) ;
  • Suh, Soo-Jhi (Department of Radiology, Keimyung University School of Medicine)
  • 발행 : 1989.06.01

초록

Three-dimension paraffin compensator was designed to construct the tissue equivalent compensator for irregular body contours and obiliques beam incidence. The ratio of compensator thickness to tissue deficit was depended on field size, depth and air gap because the scattered dose loss. The ratio of compensator-tissue was optimized 0.79, 0.73, 0.61 and 0.56 in 6MV x-rays as function of field size $4{\times}4$, $10{\times}10$, $20{\times}20$ and $30{\times}30cm^2$ respectively. in our study. Using this tissue equivalent compensator, it can be got 2% difference of dose at same mid-plane in phantom study.