Calculation of Tissue-Air Ratios(TAR) in Irregularly shaped Field for Co-60 Gamma Radiation

CO-60 감마선에 대한 부정형조사면의 조직공중선량비 (TAR) 계산

  • Ji Young-Hoon (Dept. of Therapeutic Radiology, Kangnam General Hospital)
  • Published : 1989.05.01

Abstract

In order to calculate the dose on each interest point in five types of irregularly shaped fields used commonly in radiotherapy, the tissue-air ratios (TAR) in these fields for Go-60 gamma radiation were calculated using the newly devised SAR-chart. The TARs calculated from newly method of using the SAR-chart, computer method and approximation method at the interest point were compared to the TARs obtained from measurement. The result are as follows; In case of the interest points on central axis the calculated TARs in irregularly shaped fields by the above mentioned methods were well agreed within the error of $1\%$, whereas for the interest points on off-axis the calculated TARs were resulted in the maximum errors of $2.4\%,\;2.3\%$ and $8.8\%$ respectively. From these results, the accuracy of calculation method of using the SAR-chart was comfirmed.

Keywords