Measurement of Radon and its Daughters Concentration in Air

공기중 라돈 및 라돈 자핵종의 농도 측정

  • Park, Y.W. (Korea Advanced Energy Research Institute) ;
  • Ha, C.W. (Korea Advanced Energy Research Institute) ;
  • Ro, S.G. (Korea Advanced Energy Research Institute)
  • Published : 1989.12.20

Abstract

A Lucas cell was established and calibrated by using the double layer tube standard radon source. The calibration factors were 0.031$\pm$0.002 (pCi/l)/(cph/Cell) at room temperature, and 0.029$\pm$0.001 (pCi/l)/(cph/Cell) at $50^{\circ}C$. Radon and its daughters concentrations were measured in a room air for the demonstrating purpose. The concentrations of 222 Rn, $^{218}Po,\;224\;Pb,\;and\;^{214}Bi$ were 0.87, 0.53, 0.35 and 0.26 pCi/l. The total eqilibrium factor was around 0.40 and the WL is $3.33{\times}10^{-3}$, resulting in 30 mrem/yr at this place.

이중관 형태의 라돈 표준선원을 제작하여 루카스셀의 계수치 (cph)를 라돈농도로 나타내기 위한 환산인자를 측정한 결과 라돈 표준선원을 실온으로 하였을 때는 0.031$\pm$0.001 (pCi/l)/(cph/Cell)였다. 사무실 내에서 라돈과 라돈 자핵종의 농도를 측정한 값은 $^{222}Rn,\;{\rightarrow}^{218}Po\;{\rightarrow}^{214}Pb,\;{\rightarrow}^{214}Bi$ 의 평균 농도가 각각 0.87, 0.53, 0.35, 0.26 pCi/l 였다. 이때 전체 방사평형인자와 WL의 평균 값은 각각 0.40, 3.33${\times}10^{-3}$ 이며 측정 지점에서 연간 피폭되는 방사선량으로 환산하면 약 30 mren 이다.