Isolated Working Canine Heart Perfusion Apparatus for Evaluation of Myocardial Protection Methods

심장기능 평가를 위한 견 적출심장 관류장치의 설계

  • Published : 1988.04.01

Abstract

An in vitro model providing with a recirculating perfusion apparatus using an isolated canine heart and its autogenous blood, which was prepared for study of myocardial protection method. This apparatus was easily used by quick connect system and maintained well heart function for about 2 hours. The Langendorff perfusion was initiated for a 10 minute period by introducing perfusate at 37` into the aorta from aortic reservoir located 100 cm above the heart. The isolated perfused working canine heart model was a left heart preparation in which oxygenated perfusion medium [at 37K] entered the cannulated left atrium at a constant flow rate [900ml/ min] under 20 mmHg overflow system and was spontaneously ejected[no electrical pacing] via an cannula against a hydrostatic pressure of 80 cm H2O. During this working period, various indices of cardiac function were measured. The cardiac functions were stable for over 2 hours with perfusion of Krebs-Henseleit solution and autologous blood[1:1] mixture in volume and maintained heart rate ]]3-122/bpm peak systolic pressure 109-113 mmHg, cardiac output 900 ml / min and left atrial mean pressure 8-9 mmHg. In this model, the efficiency of myocardia] protection could be easily measured by means of functional, enzymatic, biochemical and ultrastructural assessment. And also, we believe this model to be a useful assessment screening model of recovery state after long duration of myocardial preservation of donor heart without difficult transplantation procedures.