Contamination of the 6MV Linear Accelerator Photon Beam by Electrons

6MV 전자 선형 가속기 광자선의 전자오염

  • Yoo, Meong-Jin (Department of Therapeutic Radiology, Pusan National University Hospital) ;
  • Kim, Dong-Won (Department of Therapeutic Radiology, Pusan National University Hospital) ;
  • Chung, Woon-Hyuk (Department of Physics, Pusan National University)
  • 유명진 (부산의대 부속병원 치료방사선과) ;
  • 김동원 (부산의대 부속병원 치료방사선과) ;
  • 정운혁 (부산대학교 자연과학대학 물리학과)
  • Published : 1988.06.20

Abstract

The 6 MV photon beam of a linear accelerator (Mevatron 67) was studied for electron contamination. The surface dose, attributable almost entirely to contamination electrons, has a linear dependence on field width for square fields and an inverse square dependence on distance from the bottom of the fixed head assembly. Build-up and surface dose measurements were taken with and without an acrylic blocking tray in place. Further measurements were made with a copper filter designed to reduce secondary electrons emitted by photon interactions with the acrylic tray. The results are discussed in relation to skin sparing effect for radiation therapy patients. To achieve the maximum skin sparing effect, the selection of the optimum SSD and TSD is needed.

부산의대 부속 병원의 전자선형가속기 (Mevatron 67) 장치를 이용 하여 표면선량의 원인이 되는 광자선의 전자오염에 대한 연구를 하였다. 표면선량은 조사야의 크기에 따라 증가하는 경향을 나타내었다. 빔내에 트레이를 설치했을 때와 트레이를 제거했을때 표면선량의 변화와 전자오염정도를 관찰하였다. 전자오염을 감소시키기 위해서 구리 필터를 트레이 밑에 부착하여 그 효과를 분석하였다. 아울러 거리를 변화 시켜가며 선량의 변화를 관찰하여 최대의 피부보호효과를 유지할 수 있는 최적의 SSD와 TSD를 구하였다. 이런 모든 결과들은 방사선 치료 환자의 피부보호에 이용될 것이다.