Accuracy of Dose Estimation in High Dose Rate Intracavitary Radiotherapy of Carcinoma of the Uterine Cervix

자궁경부암 고선량율 강내치료의 치료선량 정확도에 관한 연구

  • Huh, Seung-Jae (Department of Radiation Therapy, Soon Chun Hyang University Hospital) ;
  • Ha, Sung-Whan (Department of Therapeutic Radiology, College of Medicine, Seoul National University) ;
  • Chai, Kyu-Young (Department of Therapeutic Radiology, College of Medicine, Seoul National University)
  • 허승재 (순천향의대 의학부 치료방사선과학교실) ;
  • 하성환 (서울대학교 의과대학 치료방사선과학교실) ;
  • 채규영 (서울대학교 의과대학 치료방사선과학교실)
  • Published : 1987.12.01

Abstract

In brachytherapy of uterine conical cancer using a high dose rate remote afterloading system, it is of prime importance to deliver a accurate dose in each fractionated treatment by minimizing the difference between the pre-treatment planned and post-treatment calculated doses. The post-treatment calculated point A dose was not much different from the pretreatment planned dose (500 cGy). The $average{\pm}standard$ deviation was $500\pm18cGy$ and 84 percent of 82 intracavitary radiotherapy was within the range of $500\pm25cGy$.

자궁경부암의 고선량율 강내치료의 매 분할 치료시마다 강내 applicator의 위치 차이로 인하여 조사 선량이 치료전 계획된 선량과 차이가 있을 수 있다. 저자는 자궁경부암 환자의 강내치료시 치료전 계획된 A점 선량 500 cGy와 치료직후에 촬영한 전후 및 측면 선 사진을 이용하여 계산된 A점 선량을 비교한 결과, 82예의 강내치료에서 치료 후 계산된 조사 선량과 계획 선량은 $500\pm18 cGy$이었으며 $84\%$에서 $500\pm25 cGy$의 범위에 포함되었다. 이러한 결과에서 계획 선량과 치료 후 계산된 조사 선량사이에 비교적 높은 일치율을 확인할 수 있었다.