A Measurement of Kerma and Absorbed Dose in Photon Fields

Photon Beam에 대한 Kerma와 흡수선량의 측정

  • Kim, Sung-Hee (Dept. of Physics, College of Science, Ewha Womans University) ;
  • Shin, Seung-Aea (Dept. of Physics, College of Science, Ewha Womans University) ;
  • Chu, Sung-Sil (Dept. of Radiology, College of Medicine, Yonsei University)
  • 김성희 (이화여자대학교 자연과학대학 물리학과) ;
  • 신승애 (이화여자대학교 자연과학대학 물리학과) ;
  • 추성실 (연세대학교 의과대학 치료방사선학교실)
  • Published : 1986.06.20

Abstract

Determination of the relation between the kerma(Kinetic Energy Released in Material) and the absorbed dose is one of the basic problems of dosimetry. Kerma and absorbed dose were measured for 6 MV X-ray from the high energy medical linear accelerator and $^{60}Co$ gamma-ray. The experimental results show that the absorbed dose in the transient equilibrium region practically coincide with the kerma in water and Al for $^{60}Co$. The maximum dose depths were $1.45g/cm^2$ for 6MV X-ray and $0.48g/cm^2\;for\;^{60}Co$ gamma-ray. The ratios of the absorbed dose at maximum build-up to the collision kerma at the surface, ($K^{att}$), were 0.949 for 6MV X-ray and 0.992 for $^{60}Co$ gamma-ray. No difference was found between water and Al when the standard field size was used. This results show that the dependence of $K^{att}$ on the material is very small.

Kerma와 흡수선량 사이의 관계를 결정하는 것은 선량측정의 기본적인 문제이다. 본 연구에서는 고에너지 치료용 선형가속기의 6MV X-선과 감마치료기의 $^{60}Co$에 대한 kerma와 흡수선량을 측정하였다. 본 실험결과는 $^{60}Co$감마선에 의한 물과 알루미늄의 과도 평형 영역에서의 흡수선량이 실제적으로 kerma와 일치하였으며, 최대선량 깊이는 6MV와 $^{60}Co$에서 각각 $1.45g/cm^2$$0.48g/cm^2$이었다. 최대 build-up에서의 흡수선량과 표면에서의 충돌 kerma의 비($K^{att}$)는 표준 조사면에서 물과 알루미늄에 대해, 6MV인 경우는 0.949, $^{60}Co$인 경우는 0.992이었다. 이 결과는 $K^{att}$의 물질에 대한 의존성이 매우 작음을 보여준다.

Keywords