Trouble in Source Driving System of a $^{60}Co$ Teletherapy Unit

$^{60}Co$ 치료장치의 선원 구동상의 문제점

  • Kang, Wee-Saing (Department of Therapeutic Radiolgy, College of Medicine, Seoul National University)
  • 강위생 (서울대학교 의과대학 치료방사선과학교실)
  • Published : 1986.06.20

Abstract

An asymmetry in dose profile of a $^{60}Co$ teletherapy unit was found by means of water-phantom measurement. The reason of that trouble was confirmed to be the abnormal 'ON' position of the source, which is resulted from the high friction between contiguous surfaces of the spring for driving the source to 'OFF' position. Lubrication in the spring improved the mobility a little, but was not a radical repair. The radical repair was to replace the old spring by new one. Periodic maintenance for source driving system and periodic measurement of field symmetry are required for prevention of abnormal 'ON' position of $^{60}Co$ source.

Picker C-9 $^{60}Co$원격치료장치에 의한 선량분포의 비대칭을 물팬톰에 의한 측정으로부터 발견하고 비대칭적인 선량분포의 원인이 선원의 비정상적인 'ON' 위치였음을 확인하였다. 선원이 비정상적인 위치에서 정지하게 된 원인이 선원을 'OFF'위치로 되돌리는 태엽의 일부 인접면이 접촉하게되어 마찰의 발생과 증대였음을 확인하였다. 태엽에 윤활유를 쳤을 때 선량분포의 대칭성이 다소 호전되었지만 근본적인 해결책은 태엽을 교체하는 것이다. 예방대책으로는 주기적인 선량분포의 측정에 의한 확인과 보수유지계획에 의한 관리가 수행되어야 할 것이다.