A Study on the Dose Distribution of Various Field and Penumbra Shield in the Telecobalt-60

코발트-60의 조사야(照射野) 변형(變形) 및 반음영(半陰影) 차폐(遮蔽)효과에 따른 선량분포(線量分布)에 관한 연구(硏究)

  • 김영일 (신구전문대학 방사선과) ;
  • 이혜경 (신구전문대학 방사선과)
  • Published : 1985.12.30

Abstract

This study was performed on the dose distribution of various field size and the effect of penumbra shield in the telecobalt unit. The results obtained are as follows. 1. Errors of the light and ${\gamma}-ray$ field size was below the regulation as 0.52 percentage. 2. The coefficient of field area was increased with the larger field area, and this coefficient was showed the more difference in larger SSD. 3. The rectangular field areas, which were described by level of the same percentage depth does, were decreased with the more elongation factor. At the same elongation factor, the compensating factor was decreased with the larger field size. 4. The lead block or extension collimator was able to shield r-ray exposure of outside field size from 50 to 80 percentage. 5. On the matching adjacent fields, while the gap between beam edges are contacted, that overlapped beam edges indicated up to 140 percentage, and while the gap was 1 cm, it could be reduced to 90 Percentage. The lead-libocking on the overlapped area was more effective to lower dose, as 80 percentage in this case. 6. Percentage depth dose of various trimming field sizes were increased linearlly according to area 1 perimeter size, but the center split field size did not maintain linearlly.