Preparation of a Liquid Membrance Type Ion-Selective Electrode and Its Application to the Potentiometric Titration

액체막형 구리이온 선택성 전극의 제작과 전위차적정에의 응용

  • Heung Lark Lee (Department of Chemistry, College of Natural Sciences, Kyungpook University) ;
  • Seung Tae Yang (Department of Chemistry, College of Natural Sciences, Kyungpook University)
  • 이흥낙 (경북대학교 자연과학대학 화학과) ;
  • 양승태 (경북대학교 자연과학대학 화학과)
  • Published : 1985.04.20

Abstract

Construction of a liquid membrane type of cupric ion selective electrode and its application to the potentiometric titration have been studied. A liquid ion-exchange membrance was prepared by extracting Cu(II) in aqueous solution into 1-(2-pyridylazo)-2-naphthol/nitrobenzene. A Ag/AgCl internal reference electrode was dipped into the aqueous reference solution of $1.00 {\times} 10^{-3}M\;Cu(NO_3)_2$ buffered with HAc-NaAc buffer solution, which was in contact with the nitrobenzene extract. The electrode showed the nernstian response to Cu(II) in the concentration range from $1.00{\times} 10^{-6}$ to $1.00{\times} 10^{-3}$M. The most suitable ion-exchanger concentration in the liquid membrane was $1.00{\times} 10^{-4}$M. The selectivity coefficients of the electrode for the various metal cations were investigated. The electrode was applied to the potentiometric titration of Cu(II) with EDTA.

액체막형 구리이온 선택성 전극의 제작과 전위차적정에의 응용을 연구하였다. 액체 이온교환막은 수용액속의 Cu(II)을 1-(2-pyridylazo)-2-naphthol/nitrobenzene 용액 속으로 추출하여 만들었다. 액체 이온교환막 위에 HAc-NaAc 용액으로 완충된 $1.00 {\times} 10^{-3}M\;Cu(NO_3)_2$ 용액을 채우고 Ag/AgCl 내부 기준전극을 담그었다. 이 이온선택성 전극은 $1.00 {\times} 10^{-6}$ ~ $1.00 {\times} 10^{-3}$M Cu(II) 농도 범위에서 nerstian response를 보여주었다. 가장 적절한 이온교환체의 농도는 $1.00 {\times} 10^{-4}$M이었다. 여러가지 금속이온에 대한 이온선택성 전극의 선택계수를 측정하였다. EDTA를 적정제로 하여 Cu(II)에 대한 전위차적정에 응용하였다.

Keywords

References

  1. Z. Biol. v.47 M. Cremer
  2. Methods Organic Analysis. Using Ion-Selective Electrodes T.S. Ma;S.S.M. Hassen
  3. The Principles of Ion-Selective Electrodes and of Membrane Transport E. Morf
  4. Analysis with Ion-Selective Electrodes(2nd Ed.) P.L. Bailey
  5. Ion-Selecitve Electrode Methodology A.K. Covington
  6. Analysis with Ion-Selective Electrodes J. Vesely;D. Weiss;K. Stulik;M. Stulikova(tr.)
  7. Ion-Selective Electrodes J. Koryta
  8. Pure Appl. Chem. v.48 IUPAC Recommendations for nomenclature of ion-selective electrodes
  9. Selectivity Coefficients for Ion-Selective Electrodes K. Umezawa;Y. Umezawa
  10. Science (N.Y.) v.156 J.W. Ross
  11. Pittsburg Conference on Analytical Chemistry and Applied Spectroscopy Liquid Ion Exchange Membrane Electrodes J. Ross;M. Frant
  12. Anal. Chem. v.41 J. Harrell;A. Jones;G. Choppin
  13. Trudy Mosk. Khim. Teknol lnst. v.1972 A. Gordievskii;A. Syrchenkov;S. Sauvin;V. Shterman;S. Chizhevskii;A. Zhukov
  14. Chem. Anal. (Warsaw) v.24 W. Szczepaniak;M. Ren;K. Rem
  15. Anal. Chirn. Acta v.82 W. Szczepaniak;K. Rem
  16. Ion-Selective Electrodes(3rd Ed.) V.A. Zarinskii;L.K. Spigun;G.E. Viasova;I.V. Volobueva;E.V. Pybakova;Y.A. Zolotov;E. Pungor(ed.)
  17. Anal. Chem. v.54 M.E. Meyerhoff;Y.M. Fraticelli
  18. Anal. Chem. v.56 M.A. Arrmold;M.E. Meyerhoff
  19. Electrochirn. Acta v.21 R. Smits;D.L. Massart
  20. Denki Kagaku v.43 M. Kataoka;T. Kambara
  21. Anal. Chem. v.41 K. Momokii(et al.)
  22. Talanta v.20 N. Ishibashi;H. Kohara;K. Horinochi
  23. Solvent Extraction of Nletals A.K. De;S.M. Khopkar;R.A. Chalmers
  24. Analyst v.97 D. Midgley;K. Torrance
  25. Analyst v.92 R.G. Anderson;G. Nickless
  26. H.L. Lee;M.L. Suh
  27. Zh. Fiz. Khim. v.10 B.P. Nicolsky
  28. Natl. Bur. of Standards Spec. Publ. v.314 Ion-Selective Electrodes G. Eisenman;R.A. Durst (ed.)
  29. Anal. Chem. v.41 K. Srinivasan;G.A. Rechnitz
  30. Anal. Chem. v.43 S. Lai;G.D. Christian
  31. Anal. Chem. v.46 R. Jasimski;L. Trachtemberg;D. Andrychuk
  32. J. Amer. Chem. Soc. v.59 J. Kielland