International Comparison of Absolute Activity Measurement of $^{133}Ba$ Solution

$^{133}Ba$ 용액의 방사능 절대측정의 국제비교

  • Park, Tae-Soon (Radiation Standards Laboratory, Korea Standards Research Institute) ;
  • Oh, Pil-Jae (Radiation Standards Laboratory, Korea Standards Research Institute) ;
  • Hwang, Sun-Tae (Radiation Standards Laboratory, Korea Standards Research Institute)
  • 박태순 (한국표준연구소.방사선표준연구실) ;
  • 오필제 (한국표준연구소.방사선표준연구실) ;
  • 황선태 (한국표준연구소.방사선표준연구실)
  • Published : 1985.12.20

Abstract

The activity measurement of a solution of $^{133}Ba$ which is an electron capture nuclide was carried out by the ${\beta}-{\gamma}$ coincidence method. The counting rates at the ${\beta}-,\;{\gamma}-$, and coincidence-channels were measured using a $4{\pi}$ proportional counter and two NaI(Tl) scintillation detectors. The specific activity of the solution calculated by the efficiency extrapolation was $(1151.01{\pm}2.99)kBqg^{-1}$ at the reference time(00h UT, 03-15-84). According to an international comparison of activity measurements organized by the Bureau International des Poids et Mesures, this result showed the difference of 0.94% to the mean value derived from the comparison.

전자포획 핵종인 $^{133}Ba$ 용액의 방사능 측정이 ${\beta}-{\gamma}$ 동시계수 방법에 의하여 수행되었다. $4{\pi}$ 비례계수기와 두개의 NaI(Tl) 섬광검출기를 사용하여 베타, 감마 및 동시계수 채널에서 계수율을 측정하였다. 효율외삽법으로 산출된 용액의 비 방사능은 기준시간(1984년 3월 15일 0시, UT)에 대하여 $(1151.01{\pm})2.99)\;kBq{\cdot}g^{-1}$이었다. 국제도량형국이 주관한 방사능 측정의 국제비교에 의하면, 이 결과는 그 비교에서 얻어진 평균값에 대하여 약 0.84%의 차이를 나타냈다.

Keywords