A Study on Ultrasound Pulsed Doppler Systems for Sending the Blood Flow

혈류 진단을 위한 초음파 펄스 도플러 시스템에 관한 연구

  • Kim, Seong-Ryul (Dept. of Electrical Eng. and Electronic Eng., Korea advanced Institute of Science and Technology) ;
  • Kim, Jin-Ha (Dept. of Electrical Eng. and Electronic Eng., Korea advanced Institute of Science and Technology) ;
  • Park, Song-Bae
  • 김성률 (한국과학기술원 전기 및 전자공학과) ;
  • 김진하 (한국과학기술원 전기 및 전자공학과) ;
  • 박송배
  • Published : 1984.09.01

Abstract

In the conventional pulsed doppler system, gated CW is used to excite a ultrasonic transducer so that a group of linear RF amplifiers are required to excite a ultrasonic array transducer in the scanning pulsed doppler system. A pulsed doppler system without linear high voltage RF amplifiers, which excites the transducer impulsively, is studied theoretically and experimentally. In this paper, an experimental 8-channel pulsed doppler system is implemented, which uses quadrature defection to detect the direction of motion and can compensate the attenuation effect. The designed pulsed doppler system shows the possibility of real time multichannel doppler flow meter.

보통의 초음파 도플러 시스템은 게이트된 연속파(gutted CW)를 사용하기 때문에 초음파 B스캔너와 같이 스캔할 수 있는 초음파 도플러 진단기는 많은 선형 고주파 증폭기를 필요로 한다. 제작이 어려운 고압 고주파 증폭기를 사용하지 않고 초음파 소자를 직접 임펄스로써 구동하는 펄스 도플러 시스템이 이론적으로 해석되었으며, 실험적으로 연구되었다. 설계된 시스템은 8채널 기능을 갖고, 직각 검파(quodrature detection)를 사용하여 목표물의 운동방향을 알 수 있도록 하였으며, 감쇄영향을 보상할 구 있도록 하였다. 연구된 시스템은 주파수 축상에서 실시간으로 혈류진단을 할 수 있는 다중 채널 진단기의 가능성을 보이고 있다.