A Study on Dose Calculation in Intracavitary Radiotherapy of the Carcinoma of the Uterine Cervix with TAO Applicator

TAO Applicator를 이용한 자궁경암 강내조사시의 선양계산에 관한 고찰

  • Kim, Chul-Soo (Department of Radiology, Hanyang University Hospital) ;
  • Kim, Jung-Jin (Department of Radiology, Hanyang University Hospital)
  • 김철수 (한양대학교 의과대학 방사선과학교실) ;
  • 김정진 (한양대학교 의과대학 방사선과학교실)
  • Published : 1984.06.01

Abstract

Various methods are available for determination of exposure time in intracavitary radiotherapy of the carcinoma of the uterine cervix. To determine the accuracy of dose calculation with isodose curve for TAO applicator, comparison with results calculated by computer for radiotherapy treatment Planning was done in 24 procedures done in 12 consecutive patients with the carcinoma of the uterine cervix from May to December, 1983. The results are as follows: 1. The average dose rate Per hour of Point A was 87.70 rad, being 89.91 rad ana 85.49 rad in left and right, respectively. 2. The average percentage of dose rate of point A calculated by isodose curve method over that by computer was $101.28\%$ and the difference was less than $5\%$ in 17 Procedures and over $10\%$ in only 3 procedures. 3. The average percentage in case of point B was $108.67\%$. In conclusion, in most cases the difference was less than 200 rad for point A and less than 100 rad for point B during 2 courses of intracavitary radiotherapy. And so the dose rate calculation with isodose curve for TAO applicator is comparatively accurate.