Change of Dose Distribution on the Beam Axis of 60Co γ Ray and 10MV X-Ray with Part Thickness

치료부위(治療部位)두께에 따른 Co-60 γ선(線)과 10MV X선(線)의 선축상(線軸上) 선량분포(線量分布)의 변화(變化)

  • Kang, Wee Saing (Department of Radiology, College of Medicine, Seoul National University) ;
  • Koh, Kyoung Hwan (Department of Radiology, College of Medicine, Seoul National University) ;
  • Ha, Sung Whan (Department of Radiology, College of Medicine, Seoul National University) ;
  • Park, Charn Il (Department of Radiology, College of Medicine, Seoul National University)
  • 강위생 (서울대학교 의과대학 방사선과학교실) ;
  • 고경환 (서울대학교 의과대학 방사선과학교실) ;
  • 하성환 (서울대학교 의과대학 방사선과학교실) ;
  • 박찬일 (서울대학교 의과대학 방사선과학교실)
  • Published : 1983.10.30

Abstract

The thickness of the part being irradiated is finite. Percent depth dose tables being used routinely are generally obtained from dosimetry in a phantom much thickner than usual patient. At or close to exit surface, the dose should be less than that obtained from the percent depth dose tables, because of insufficient volume for backscattering. To know the difference between the true absorbed dose and the dose obtained from percent depth dose table, the doses at or close to the exit surface were measured with plate type ionization chamber with volume of 0.5ml. The results are as follows; 1. In the case of $^{60}Co$, percent depth dose at a given depth increases with underlying phantom thickness up to the 5cm. 2. In the case of $^{60}Co$, the dose correction factor at exit surface which is less than 1, increases with part thickness and decreases with field size. 3. Exposure time may not be corrected when the part above 10cm in thickness is treated by $^{60}Co$. 4. In the case of 10MV x-ray, the dose correction factor is nearly 1 and constant for the underlying phantom thickness and field size, so the correction of monitor unit is not necessary for part thickness.

Co ${\gamma}$선과 10MV x선으로 폴리스티렌 팬톰에 조사하여 이온함으로 배면과 배면근처의 선량을 측정한 결과는 아래와 같다. 1. 측정점 이하의 팬톰의 두께가 증가함에 따라 선량율과 선량보정계수가 증가하였으며 그 두께가 20cm 이상인 경우에는 일정하였다. 선량보정계수는 1 이하였다. 2. 조사면의 크기가 클수록, 측정점의 깊이가 깊을 수록 선량보정계수는 감소하였다. 3. 선량에 후방산란선량의 기여도는 조사면이 클수록, 깊이가 얕을수록, 그 깊이 이하의 물질의 두께가 두꺼울수록 증가하였다. 4. 10MV x선의 경우에도 선량보정계수는 같은 유형으로 변하기는 했지만 그 값이 0.995~1.000이었다. 5. $^{60}Co$에 경우에 치료부위의 두께가 10cm 이상이면 조사시간을 보정하지 않아도 좋을 것이다.