A Study on the Adaptability and Cytotoxicity of Porcelain Bonded Ni-Cr Alloy

도재소부용 Ni-Cr합금의 적합성과 세포독성에 관한 연구

  • Choi, Boo-Byung (Dept. of Prosthodontics, College of Dentistry, Kyung Hee University) ;
  • Shin, Myung-Chull (Dept. of Prosthodontics, College of Dentistry, Kyung Hee University)
  • Published : 1982.12.01

Abstract

In order to investigate the internal adaptability and cytotoxicity for porcelain fused to metal Ni-Cr alloy, two commercial alloys (Generic Gold-R alloy, Aalba Dent- Valloy)were employed and compared with new developed beryllium free Ni-Cr alloy (KIST-K alloy). After cementation of each crown to abutment die before and after ceramic bonding respectively, dies were mounted to epoxy resin and the internal space was measured between crown and die of each point on sagittal section. The results were as follows: 1) The results show that the space between the crown and the die was largest at the incisal edge, followed by the shoulder or the bevel and chamfer. The least was to be in the axial surface. 2) There were no significant difference in the adaptability of the compared alloys. 3) On day 2 of the experiment the cell multiplication was slightly inhibited by R and K alloys. However, observations taken at day 4 and 7 showed normal cell multiplication process compared to control group. 4) Alloy V exhibited the most severe cell toxicity among the alloys tested on day 2. By day 4 and 7 it showed a decrease level in toxicity, however it did not exhibit normal cell growth as compared to control group.

Keywords