A Study on the Degree of Contamination of Air and Materials in a Hospital

병원내 공기오염과 물품의 항균에 대한 상태조사

  • 차옥주 (서울대학교 보건대학원)
  • Published : 1982.04.01

Abstract

The incidence of hospital infection has been seriously increased in the general hospital in recent years. This study was performed on hospital air and materials in a General Hospital in Seoul from June to December in 1980. The results were as follows: 1. Air sampling was done in multiple strategic areas by exposing standard petridishes for 5 minutes. There was a significant difference of airborne microbe between places. ($F._{99}$ = 3.2, p < 0.01). 2. The mean colony count was 8.6$\pm$6.2 colonies / plate / 5 minutes. 3. Gram stains of colony in air sampling were Gram (+) cocci 66.5%. Gram (+) rod 18.4%, Gram (-) cocci 1.3%, Gram (-) rod 8.7% Fungus 4.5%. 4. For the evaluation of sterilization of steam sterilizer and ethylene oxide gas sterilizer, biological monitoring were done by commercial spore strip. Positive culture was obtained in 2 out of 41 tests on 3 steam sterilizers, and in 3 out of 13 tests on ethylene oxide gas sterilizer. 5. Product sampling and culture were done for 2 kinds of disinfectants and 30 sets of various operation package or dressing materials. Positive culture was obtained in one disinfectants.