관전압(管電壓)이 화상(畵像)에 미치는 영향(影響)
A Study on the Image Effect of Tube Voltage
- Kang, Hong-Seok (Dept. of Radiotechnology, Junior College of Public Health and Medical Technology, Korea University) ;
- Kim, Chang-Kyun (Dept. of Radiotechnology, Junior College of Public Health and Medical Technology, Korea University) ;
- Huh, Joon (Dept. of Radiotechnology, Junior College of Public Health and Medical Technology, Korea University)
- 발행 : 1980.07.18
초록
To know the effect of tube voltage to X-ray film image, authors made an experimental study on the exponential value of applied voltage and the effect of scattered ray in photographic contrast, and obtained the results as follows: 1. The exponential value of tube voltage was under the control of the applied voltage, the kinds of screens and grids, and its existence. The range of its value was from 1.03 to 5.3. 2. The efficiency of X-ray production was directly proportional to the applied voltage, and to oftain the same density, the tube current(mAs) was inversely proportional to applied voltage 3. The production of scattered rays was in proportion to the tube voltage, and the photographic contrast was mainly influenced by the scattered rays.