A study on the flume for a current meter rating

유속계 검정용수로에 관한 연구

  • 정준석 (본회이사·한양대학교공과대학) ;
  • 박정응 (경기공업전문학교)
  • Published : 1973.12.01

Abstract

The coefficient of the current meter generally determined by the maker Its coefficient is subject to being changed with time. Therefore the coefficient of the current meter has to be checked up before it is ready to be used Such an inspection is termed a current meter rating The current meter equipped an electronic apparatus and all the others are to be rated in a rating flume. The price current meter which is most widely used for measuring flow velocities ranging between 0.3m/sec and 3.5m/sec has been used in this study. The length of the flume and the optimum range of the rating in the cross section are determined in the range of 20∼120cm deep, 50∼160cm wide of the flume. In this study, the 23 different kinds of the current meter rating enabled us to determine the constants a and b of the following equation. V=an+b(m/sec) where, n is number of revolution per second(n=N/T) V is velocity(v=D/T) The above constant can be determined by the least squares method and plotting, using the velocity(V=D/T) and the number of revolution per second(n=N/T) obtained from the running distance(D), time(T), the number of revolutin(N), and the running number(m). From the experiments the following conclusions are drawn: 1) The rating flume is large enough if the flume is 110∼120cm deep, and 40∼50m long. 2) The optimum depth for rating of a current meter is in the range of h=40∼50cm.

Keywords