A Study on the Intensity Measurng of Ultrasonics by Added Two Sensive point of Thermocouple

이감온점 열전대에 의한 초음파의 강도측정에 관한 연구

  • Published : 1971.12.01

Abstract

This paper describes concerning the method of measurement for the intensity of ultrasonics in liquid. The probe which is added two sensive point of thermocople which is surrounded by the absorbed material of sonic wave on the both side of the refleftite plate have measured the intensity of ultrasonics indirectly by the difference of the temperature of two sensive point which had raised the temperature by the sonic wave and its reflective wave which transmitted the ultrasonic energy. This minify the influence from the temparature of liquid and time constant. Consequently, this is the basic industrial method which is with in the bounds of possibility on the simple measurement of the intensity of the local ultrasonics concerning direction of its propagation.

본 연구는 액체내의 초음파강도측정법에 관한 것으로 반사판의 양측에 흡음특질로 포위된 thermocouple의 감온점을 첨부한 구조를 가진 probe에 직접 받은 음파와 반사판을 투과한 음파에 의하여 상승된 감온점의 온도차를 검출하여 초음파강도를 온정하였다. 이것은 액온의 영향과 time constant를 적게하고 초음파 진행을 고려한 국소 강도를 간편하게 측정할 수 있는 공업용 계기제작을 위한 기초연구로서 시도한 것이다.