Abstract

The purpose of this study is to develop the auxiliary rehabilitation device for a patient of spinal curvature. The adolescent diopathic scoliosis(AIS) must be treated by rehabilitation brace if Cobb angle is $20^{\circ}{\sim}40^{\circ}$. The rehabilitation brace is consist of 4 vest and 2 hinge parts(hinge and couple bar) that give a force to the ribs. But thin and light hinge parts for young patient failed easily because of unusual movement of the upper body. We studied optimum design and structural analysis of hinge parts when it distorted by tensile and bending force. The specimen of hinge parts were tested to evaluate the failure strength. And we attached circuits with memory and sensor detecting pressure and displacement to recoding stress in hinge parts. These data are used to alarm the patient to stop unusual movement and understand the load history.