Development of angiographic PIV and performance evaluation

Angiographic PIV기법 개발 및 성능 평가

  • 정성용 (포항공과대학교 대학원) ;
  • 김국배 (한국과학기술연구원) ;
  • 임남윤 (포항공과대학교 기계공학과) ;
  • 이상준 (포항공과대학교 기계공학과)
  • Published : 2008.03.26

Abstract

To diagnose vascular diseases from the viewpoint of hemodynamics, detailed quantitative hemodynamic information of related blood flows with a high spatial resolution of tens micrometer and a high temporal resolution in the order of millisecond were required. For investigating the hemodynamic phenomena of vascular circulatory diseases, a new diagnosing technique combining a medical radiography and PIV method was newly developed. This technique consisted of a medical X-ray tube, an X-ray CCD camera, a shutter module for double pulse-type X-ray source, and a synchronizer. Through several preliminary tests, the feasibility of the angiographic PIV technique was verified. For in-vivo measurements to real blood flows, we developed tracer microcapsules which were optimized to this system by encapsulating a contrast material of iodine with a matrix material of PVA (polyvinylpyrrolidone). In near future, the angiographic PIV technique would be useful for diagnosing hemodynamic phenomena of vascular diseases and for their early detection.

Keywords