Optimum electrode selection for measuring the abdominal pressure using bio-impedance method
비침습적 복압 측정을 위한 생체 임피던스 전극의 최적 위치 선정
- Published : 2007.04.27
Abstract
In this study, we determined the optimum electrode pair for measuring the abdominal pressure using bio-impedance method. Because impedance changes differ from a weight, a height, contractile force, volume of muscle and blood other or whatever of individuals, it was quantified using values of impedance change, correlation coefficient and SNR. Our results showed the optimum electrode pair (1, 9) which could detect impedance changes due to an increase of the intensity of the abdominal pressure. The correlation coefficient and quadratic function between the RMS values of EMG and the impedance changes were 0.87 and