Optimum electrode selection for measuring the abdominal pressure using bio-impedance method

비침습적 복압 측정을 위한 생체 임피던스 전극의 최적 위치 선정

  • 안양수 (전북대학교 대학원 전자정보공학부) ;
  • 김거식 (전북대학교 대학원 전자정보공학부) ;
  • 송철규 (전북대학교 전자정보공학부)
  • Published : 2007.04.27

Abstract

In this study, we determined the optimum electrode pair for measuring the abdominal pressure using bio-impedance method. Because impedance changes differ from a weight, a height, contractile force, volume of muscle and blood other or whatever of individuals, it was quantified using values of impedance change, correlation coefficient and SNR. Our results showed the optimum electrode pair (1, 9) which could detect impedance changes due to an increase of the intensity of the abdominal pressure. The correlation coefficient and quadratic function between the RMS values of EMG and the impedance changes were 0.87 and $y=0.0014x^2$+0.0620x+0.6958, respectively. It demonstrated that the abdominal pressure could be measured non-invasively and simply using bio-impedance method. We propose that this optimum electrode configuration would be useful for future studies involving the convenient measurement of abdominal pressure by ambulatory urodynamics monitoring study.