Realtime Wireless Monitoring of Abnormal ST in ECG Using PC Based System

  • Jeong, Gu-Young (Department of Mechatronics Engineering, Chonbuk National University) ;
  • Yu, Kee-Ho (Division of Mechanical and Aerospace System Engineering, Chonbuk National University) ;
  • Kim, Nam-Gyun (Division of Bionics and Bioinformatics, Chonbuk National University) ;
  • Inooka, Hikaru (Graduate School of Information Sciences, Tohoku University)
  • Published : 2004.08.25

Abstract

The ST-segment that the beginning part of T wave is the important diagnostic parameter to finding myocardial ischemia. Abnormal ST appears in two types. One is the level change, and the other is the pattern change. In this paper, we describe the monitoring of abnormal ST using PC based system. Hardware of this system consists of transmitter, receiver and PC. The function of transmitter is measuring ECG in three channels which are selected manually and transmitting the data to receiver by digital radio way. Connection with receiver and PC is by RS232C, and the data received on the PC is analyzed automatically by ECG analysis algorithm and saved to file. In the algorithm part for detecting abnormal ST, ST-segments are approximated by a polynomial. This method can detect all of the deviation and pattern change of ST-segment regardless the change in the heart rate or sampling rate. To gain algorithm reliability, the method rejects distorted polynomial approximation by calculation the difference between the approximated ST-segment and original ST-segment. In pre-signal processing, the wavelet transformation separates high frequency bands including QRS complex from the original ECG. Consequently, the process improves the performance of detecting each feature points.

Keywords