The Measurement of Soot Particle Temperatures Using a Ratio Pyrometry

Ratio Pyrometry를 이용한 매연입자 온도 계측에 대한 고찰

  • 남연우 (단국대학교 대학원 기계공학과) ;
  • 이원남 (단국대학교 기계공학과) ;
  • 이춘범 (자동차부품연구원 환경부품연구센터)
  • Published : 2004.11.26

Abstract

The ratio pyrometry has been investigated experimentally for the measurement of soot particle temperatures in a diffusion flame. A tungsten lamp calibration system was constructed and used in order to calibrate the ratio pyrometry and two-color pyrometry using a KL-factor method. Once the ratio pyrometry is properly calibrated, temperatures measured using a ratio pyrometry were virtually identical to those obtained from a KL-factor method. The effect of soot volume fraction on temperature measurement was almost negligible, and therefore, the ratio pyrometry could provide the useful temperature information of sooting flames. The potential application of a ratio pyrometry to a 2-D temperature measurement without sacrificing the accuracy was demonstrated.

Keywords