Comparison of normal and replantation digital blood flow using photo-plethysmography

Photo-Plethysmography를 이용한 정상과 재접합 수지 혈류량의 비교

  • 남기창 (연세대학교 의과대학 의학공학교실) ;
  • 김성우 (연세대학교 대학원 생체공학협동과정) ;
  • 나동균 (연세대학교 의과대학 성형외과학교실) ;
  • 김덕원 (연세대학교 의과대학 의학공학교실)
  • Published : 2004.11.12

Abstract

Up to the present, digital replantation patients has appealed different symptoms due to blood circulatory failure. But, the level of blood circulatory failure has been evaluated only by clinical symptoms, or angiography. According to the cases of digit replantation is increasing, then objective evaluation methods of the level of blood circulation failure is needed other than patient's subjective symptoms and complaints. Although angiography, doppler, electromagnetic flowmeter, laser blood flowmeter, mechanical blood flowmeter has been used for the evaluation of the blood circulatory failure, the result was affected by time, place, surrounding temperature, patient's body temperature, and even emotion. Therefore, it is pointed out with lack of availability, feasibility and reproducibility. Thus, we compared digital blood flow of dominant hand to non dominant hand, and replanted fingers to opposite normal fingers from developed photo-plethysmography. The average digital blood flow showed no difference in normal digits each other, but, replanted digits showed average of 53% (9 - 100 %) compare to opposit normal digits. As it measure relative blood flow for circulatory failure of tissue such as fingers and toes more sensitively, reliably. In conclusion, it is expected that photo- plethysmography will be very useful for diagnosis, curative effect, prognosis of blood circulatory failure in digital replantation patient.