한국정밀공학회:학술대회논문집 (Proceedings of the Korean Society of Precision Engineering Conference)
- 한국정밀공학회 2004년도 추계학술대회 논문집
- /
- Pages.1248-1251
- /
- 2004
- /
- 2005-8446(pISSN)
족관절 근위부 골교정용 기능성 체외고정장치 개발
Development of a Functional External Fixator System for Bone Deformity near Joints in Legs
초록
The functional external fixator system for bone deformity near joints in legs using the worm gear was developed for curing the difference angles in fracture bone and the lengthening bar for curing the difference length in fracture bone. Both experiments and FE analysis were performed to compare the elastic stiffness in several loading modes and to improve the functional external fixator system for bone deformity near joints in legs. The FE model using the compressive and bending FE analysis was applied the FE analysis due to the angle differences. The results show that the compressive stiffness value in experiment was 175.43N/mm, the bending stiffness value in experiment was 259.74N/mm, compressive stiffness value in FEM was 188.67N/mm, bending stiffness value in FEA was 285.71N/mm. The errors between experiments and FEA were less than 10%. The maximum stress (157MPa) to the angle of clamp was lower than the yield stress (176.4MPa) of SUS316L. The stiffnesses in both axial compressive and bending of the new fixator are about 2 times higher than other products except EBI (2003).