Development of in-situ Analysis System for Radwaste Glass Using Laser Induced Breakdown Spectroscopy
레이저유도 플라즈마분광법을 이용한 방사성폐기물 유리의 현장분석 시스템 개발
Abstract
Laser Induced Breakdown Spectroscopy(LIBS) system is being developed as an in-situ analysis system for the radioactive waste glass in the cold crucible melter. In order to complete the LIBS system, a spectrometer, a detector, and a laser were structured. An ESA 3000 (LLA Instruments GmbH, Germany) including a calibrated Kodak KAF-1001 CCD detector was selected as the spectrometer. A Q-switched Nd-YAG Brilliant(Quantel, France) laser was selected as an energy source. As the first research stage, the excitation temperatures of Fe(I) as a function of the detector's delay intervals(500, 1000, 1500, 2000ns) were evaluated using the Einstein-Boltzmann equation. The optimized excitation temperature of Fe (I) was 7820k at the delay time of 1500㎱ using the 532nm Nd-YAG laser pulse. This LIBS system will be optimized under the real environment vitrification facility in the near future and then used to be in-situ analyzed the glass compositions in the melter qualitatively.
방사성폐기물 유리화 공전 중 용융로 내에서 유리시료를 직접분석 하기 위한 레이저유도 플라즈마분광법(LIBS: Laser Induced Breakdown Spectroscopy)을 개발 중이다. LIBS 시스템을 구성하기 위하여 분광기, 검출기, 레이저 등의 장비들을 구축하였다. 분광기는 CCD(charge coupled device)가 보정되어 내장되어있는 ESA 3000을 레이저는 Q-switched Nd-YAG인 Brilliant로 구매하여 분석시스템을 구축하였다. 구축된 분석시스템 분야별 성능들을 확인하였으며 최적화 연구를 수행하였다. 첫 번째 단계로 Fe가 함유된 시료에 레이저를 주사하였을 때 발광스펙트럼을 측정하였으며 검출기의 지연시간을 변화시켰을 때의 발광스펙트럼의 특성과 이를 이용한 여기온도를 Einstein-Boltzmann 식을 이용하여 계산하였다. 시료에 532nm Nd-YAG 레이저를 주사하고 검출기의 지연시간을 500, 1000, 1500, 2000ns로 변화시켰을 때의 분광선의 intensity 및 여기온도 변화를 분석하였다. 그 결과 검출시간이 1500㎱ 일 때 여기온도는 7820k로 가장 최적의 상태를 확인하였다. 향후 이 시스템은 유리화 실증시설에 적용되어 용융로 외부로의 유리시료 이송 없이 현장에서 유리성분들의 정량분석을 수행할 예정이다.
Keywords