Applications of ICP-MS to the Determination of Trace Np and Pu in Environmental Samples with Extraction Chromatography

  • Published : 2004.02.01

Abstract

An unique extraction chromatographic column (TOA: Tri-n-octylamine on Teflon powder) with a two-stage sample loading was prepared to separate Np and Pu from the environmental matrix. Np and Pu were efficiently retained in 4 M $HNO_3$ medium on the column and easily eluted with 0.02 M oxalic acid in 0.16 M HNO$_3$at $95^{\circ}C$. The separated solutions were free from most of the matrix elements and were aspirated into the ICP-MS directly. The decontamination factor for 238U is more than 104. The instrumental detection limit for 237Np was 0.46 pg mL-l ($1.2{\times}10_{-5}$Bq mL-l), and for 239Pu was 0.48 pg mL-l ($1.1{\times}10_{-3}$ Bq mL-l). The feasibility for the determination of both elements was proved by analysing IAEA-135 reference samples, the measured values agreed with the recommended reference value.