LOC 형광검출 소자를 위한 광 다이오드의 제작 및 특성 평가

Development of Photo-diode for LOC fluorescence detector

  • 김주환 (한국과학기술연구원 마이크로시스템 연구센터) ;
  • 신경식 (한국과학기술연구원 마이크로시스템 연구센터) ;
  • 김용국 (한국과학기술연구원 마이크로시스템 연구센터) ;
  • 김상식 (고려대학교 전기공학과) ;
  • 주병권 (한국과학기술연구원 마이크로시스템 연구센터)
  • 발행 : 2003.04.19

초록

Signal detection technologies such as fluorescence, charge and electrochemical detection used in the monolithic capillary electrophoresis system to convert the biochemical reaction into the electrical signal. The fluorescence detection using photodiodes that measure fluorescence emitted from eluting molecules is widely used for the monolithic capillary electrophoresis system. In this paper, in order to fabricate a photosensor has the increased sensitivity, we investigated on the sensitivity of general type and p-i-n type diode. The p-i-n diode has higher sensitivity than photodiode. Considering these results, we fabricated p-i-n diodes on the high resistive$(4k{\Omega}{\cdot}cm)$ wafer into rectangle and finger pattern and compared internal resistance of each pattern. The internal resistance of p-i-n diode can be decreased by the application of finger pattern has parallel resistance structure from $571{\Omega}$ to $393{\Omega}$.

키워드