Proceedings of the KIEE Conference (대한전기학회:학술대회논문집)
- 2003.11c
- /
- Pages.727-730
- /
- 2003
Measurement of the pain responses induced by electric pulp test
전기 치수 자극에 의해 유발된 동통 반응의 측정
Abstract
Electric pulp test is a method to examine the vitality of dental pulp using the electrical stimulation. During the pulp test, the current stimulates intradental nerve, and makes patients painful. Some studies were accomplished to measure the responses of subjects by stimulating over the sensory threshold. In this study, we examined the time delay between pain feeling and stimulation stop in clinical situation. And we measured the activated responses(EMG in anterior belly of digastric muscle, voice, and finger span). As a result, it was verified that the minimum and maximum delay was EMG and voice, respectively. By reducing the excessive stimulus time, the unnecessary pain can be minimized using EMG that has the minimum delay.