Measurement of the pain responses induced by electric pulp test

전기 치수 자극에 의해 유발된 동통 반응의 측정

  • 안선회 (연세대학교 대학원 생체공학 협동과정) ;
  • 남기창 (연세대학교 대학원 생체공학 협동과정) ;
  • 김수찬 (연세대학교 의과대학 의학공학교실) ;
  • 김재성 (연세대학교 대학원 생체공학 협동과정) ;
  • 이승종 (연세대학교 치과대학 보존과학교실) ;
  • 김덕원 (연세대학교 의과대학 의학공학교실)
  • Published : 2003.11.21

Abstract

Electric pulp test is a method to examine the vitality of dental pulp using the electrical stimulation. During the pulp test, the current stimulates intradental nerve, and makes patients painful. Some studies were accomplished to measure the responses of subjects by stimulating over the sensory threshold. In this study, we examined the time delay between pain feeling and stimulation stop in clinical situation. And we measured the activated responses(EMG in anterior belly of digastric muscle, voice, and finger span). As a result, it was verified that the minimum and maximum delay was EMG and voice, respectively. By reducing the excessive stimulus time, the unnecessary pain can be minimized using EMG that has the minimum delay.