Emulator Circuit for a Flux Locked Loop for Detection of Magnetocardiography Signal

심자도 신호 검출을 위한 Flux Locked Loop (FLL) Emulation 회로

  • Published : 2003.07.01

Abstract

Magnetocardiography is a very weak biomagnetic field generated from the heart. Since the magnitude of the biomagnetic field is in the order of a few pico Tesla, it is measured with a superconducting quantum interference device (SQUID). SQUID is a transducer converting magnetic flux to voltage, however, its range of linear conversion is very restricted. In order to overcome the narrow dynamic range. a flux locked loop is used to feedback the output field with opposite polarity to the input field so that the total Held becomes zero. This prevents the operating point of the SQUID from moving too far away from the null point thereby escape from the linear region. In this paper, an emulator for the SQUID sensor and feedback coil is proposed. Magnetic courting between the original field and the generated field by the feedback coil is emulated by electronic circuits. By using the emulator, FLL circuits are analyzed and optimized without SQUID sensors. The emulator may be used as a test signal for multi-channel gain calibration and system maintenance.

Keywords