Proceedings of the IEEK Conference (대한전자공학회:학술대회논문집)
- 2003.07e
- /
- Pages.1863-1866
- /
- 2003
Correction of MRI Artifact due to Planar Respiratory Motion
호흡운동에 의한 MRI 아티팩트의 수정
Abstract
In this study, respiratory motion is modeled by a 2-Dimensional linear expanding-shrinking movement. According to the introduced model, respiratory motion imposes phase error, non-uniform sampling and amplitude modulation distortions on the acquired MRI data. When the motion parameters are known or can be estimated, a reconstruction algorithm based on superposition method was used to removed the MRI artifact. For the purpose of estimating unknown motion parameters, we applied the spectrum shift method to find the respiratory fluctuation function, the x directional expansion coefficient and its center, and also we used the minimum energy method to find the y directional expansion coefficient and its center. The effectiveness of this presented method is shown by Computer simulations.
Keywords