Evaluation of electron dose distribution obtained from ADAC Pinnacle system against measurement and Monte Carlo method for breast patients
- Lee, S. (Department of Radiation Oncology, College of Medicine, Ewha Womans University) ;
- Lee, R. (Department of Radiation Oncology, College of Medicine, Ewha Womans University) ;
- Park, D. (Department of Radiation Oncology, National Cancer Center) ;
- S. Suh (Department of Radiation Oncology, College of Medicine, Ewha Womans University)
- Published : 2003.09.01
Abstract
Introduction: With the development of dose calculation algorithms for electron beams, 3D RTP systerns are available for electron beam dose distribution commercially. However, no studies evaluated the accuracy of dose calculation with ADAC Pinnacle system for electron beams. So, the accuracy of the ADAC system is investigated by comparing electron dose distributions from ADAC system against the BEAMnrc/DOSXYZnrc. Methods: A total of 33 breast cancer patients treated with 6, 9, and 12MeV electrons in our institution was selected for this study. The first part of this study is to compare the dose distributions of measurement, TPS and the BEAMnrc/DOSXYZnrc code in flat water phantom at gantry zero position and for a 10
Keywords