Application of Amorphous wire to ECT(Eddy Current Testing) Probe

아몰퍼스 와이어의 ECT probe 적용에 대한 검토

  • 김영학 (부경대학교 전기공학과) ;
  • 신광호 (경성대학교 멀티미디어 공학부) ;
  • 사공건 (동아대학교 전기공학과)
  • Published : 2002.08.24

Abstract

ECT(eddy currentign testing) is very effective technique to detect a flaw within a conductor. Co-based amorphous wire was used as a sensor head. The wire has almost 0 magneto-striction and high permeability. An uniform magnetic field was applied to 1mm thick copper plate and $25{\mu}m$ thick aluminum sheet conductor using spiral typed coil The size of the coil has $40mm{\times}40mm$ outer width and $8mm{\times}8mm$ inner width. The copper plate and aluminum sheet has 0.5mm and 0.1mm wide gap, respectively. The frequency range of applied field was 100kHz-600kHz. The induced voltage difference of 2.5mV was obtained in the maximum voltage and minimum one measured across the gap of the 1mm thick conductor. In the case of aluminum sheet, 0.4mV was obtained. From this results, the effectiveness of Co-based amorphous wire was confirmed in the ECT technique.