대한전기학회:학술대회논문집 (Proceedings of the KIEE Conference)
- 대한전기학회 2002년도 합동 추계학술대회 논문집 정보 및 제어부문
- /
- Pages.418-421
- /
- 2002
전자의료기기의 전자파 적합성 평가에 관한 연구
A Study on the Electromagnetic Compatibility Evaluation of the Electronic Medical Instrumentation
- Song, C.G. (School of Electronics & Information Eng., Chonbuk National University) ;
- Kim, J.C.
- 발행 : 2002.11.30
초록
The immunity testing in this study was performed to examine the conducted immunity and radiated immunity in medical instrumentation. Experiments were performed to assess the conducted emission and radiated emission of medical instrumentation such as MRI, ultrasound system, and electro surgical unit etc. The emission testing revolves around conducted emissions (high-frequency currents) on power lines and signal cables. Emission testing also looks at the output of energy through the air (called radiated emissions) and analyzes the strength of those radio-frequency signals generated unintentionally by the electrical medical device that is under test. Finally, this study would provide significant data for furture research or electromagnetic interference and compatibility test in medical instrumentation.
키워드