Accuracy of image registration for radiation treatment planning using a brain phantom

  • Jin, Ho-Sang (Dept. of Biomedical Engineering, The Catholic University of Korea) ;
  • Suh, Tae-Suk (Dept. of Biomedical Engineering, The Catholic University of Korea) ;
  • Song, Ju-Young (Dept. of Biomedical Engineering, The Catholic University of Korea) ;
  • Juh, Ra-Hyeong (Dept. of Biomedical Engineering, The Catholic University of Korea) ;
  • Kwark, Chul-Eun (Dept. of Biomedical Engineering, The Catholic University of Korea) ;
  • Lee, Hyoung-Koo (Dept. of Biomedical Engineering, The Catholic University of Korea) ;
  • Choe, Bo-Young (Dept. of Biomedical Engineering, The Catholic University of Korea)
  • 발행 : 2002.11.01

초록

Purpose: The purposes of our study are (1) to develop a brain phantom which can be used for multimodal image registration, (2) to evaluate the accuracy of image registration with the home-made phantom. Method: A brain phantom which could be used for image registration technique of CT-MR and CT-SPECT images using chamfer matching was developed. The brain phantom was specially designed to obtain imaging dataset of CT, MR, and SPECT. The phantom had an external frame with 4 N-shaped pipes filled with acryl rods for CT, MR imaging and Pb rods for SPECT imaging. 8 acrylic pipes were inserted into the empty space of the brain phantom to be imaged for geometric evaluation of the matching. Accuracy of image fusion was assessed by the comparison between the center points of the section of N-shaped bars in the external frame and the inserted pipes of the phantom. Technique with partially transparent, mixed images using color on gray was used for visual assessment of the image registration process.

키워드