Purpose: The purpose of this study is to optimize the configuration of body RF coil composed of 4 planar subcoils for low field open MRI. Method: Our low field RE coil is composed of 4 subcoils assumed to be located at both the bottom and top sides of permanent magnet. Each subcoils has 3 main strips. The coil system has mirror inversion symmetry. First, the currents on the strips are obtained by inductance calculation and circuit analysis, Second, all the strips are divided into line strip elements across the strips, the self Inductances of line strip elements and the mutual inductances among the line strip elements are calculated, and current distributions of strip are obtained by circuit analysis, where each strip is considered as parallel combination of line strip elements. Finally all the line strip elements are segmented, magnetic field has been calculated by pseudo electric dipole radiation method, where the current elements are regarded as pseudo electric dipole radiation sources. We have performed above procedures for various configurations of RE coil. The field homogeneity is calculated in the 25 cm DSV.