Development and Performance Evaluation of the First Model of 4D CT-Scanner
- Endo, Masahiro (National Institute of Radiological Sciences, Chiba) ;
- Mori, Shinichiro (National Institute of Radiological Sciences, Chiba) ;
- Tsunoo, Takanori (National Institute of Radiological Sciences, Chiba) ;
- Kandatsu, Susumu (National Institute of Radiological Sciences, Chiba) ;
- Tanada, Shuji (National Institute of Radiological Sciences, Chiba) ;
- Aradate, Hiroshi (Toshiba Corp. Medical System Company, Otawara) ;
- Saito, Yasuo (Toshiba Corp. Medical System Company, Otawara) ;
- Miyazaki, Hiroaki (Toshiba Corp. Medical System Company, Otawara) ;
- Satoh, Kazumasa (Sony Corp. Corporate Technology Department, Tokyo) ;
- Matsusita, Satoshi (Sony Corp. Corporate Technology Department, Tokyo) ;
- Kusakabe, Masahiro (Fukui University Faculty of Engineering)
- 발행 : 2002.09.01
초록
4D CT is a dynamic volume imaging system of moving organs with an image quality comparable to conventional CT, and is realized with continuous and high-speed cone-beam CT. In order to realize 4D CT, we have developed a novel 2D detector on the basis of the present CT technology, and mounted it on the gantry frame of the state-of-the-art CT-scanner. In the present report we describe the design of the first model of 4D CT-scanner as well as the early results of performance test. The x-ray detector for the 4D CT-scanner is a discrete pixel detector in which pixel data are measured by an independent detector element. The numbers of elements are 912 (channels)