Radiochromic film dosimetry for linac-based stereotactic radiosurgery

  • Han, Seung-Hee (Department of Radiation Oncology, Kang Dong Sacred Heart) ;
  • Park, Suk-Won (Department of Radiation Oncology, Kang Dong Sacred Heart) ;
  • Oh, Do-Hoon (Department of Radiation Oncology, Kang Dong Sacred Heart)
  • 발행 : 2002.09.01

초록

In linac-based stereotactic radiosurgery, assuring the quality of the planning and delivery of external photon beam requires accurate evaluation of beam parameters, usually including output factors, tissue-phantom ratio and off-axis ratios, and measurement of actual dose distributions from simulated treatment. We're going to test the use of calibrated radio chromic film (Gafchromic film; type MD-55, Nuclear associate) using a Lumiscan 75 digitizer to measure absolute dose and relative dose distributions for linac-based radiosurgery unit Relative dose distribution of a human-style spherical acryl phantom were measured using radiochromic film and calculated by treatment planning system. The absolute dose at the sphere center was measured by radiochromic film and micro chamber (Exradin A-14, 0.009cc). What we want to demonstrate in this work, the 'well selected' radiochromic films when external photon beam are used in linac-based stereotactic radiosurgery are very accurate detector for dosimetry.