A Study on the Improvement of the Strut for the HCF - Motorized Strut & Digital Gage

HCF용 스트럿의 개선을 위한 연구 - Motorized Strut & Digital Gage

  • Published : 2001.10.01

Abstract

Two ways of improvement of the strut for the HCF was made in this paper. The strut for the existing HCF is a passive link, which results the posture of the HCF and related bones. The accuracy of the HCF depends on the accuracy of the strut length. A "digital gage" was proposed to increase the accuracy of the strut by presenting the measuring result as figures in manual mode operation. And a "motorized strut system" was designed for the automated HCF operation. A strut was equipped with a motor, "motorized strut" can be operated manually and automatically. In automatic mode, the HCF operating data is generated by the HCF schedule package in PC and is downloaded to the "motorized strut system" controller. By these two improvements, changes in orthopedic equipments like HCF and other Ilizarov fixators are expected.